Tytek Medical Recalls TM-317 PneumoDart-Pneumothorax Needle

Tytek Medical is recalling its PneumoDart-Pneumothorax Needle due to the risk of blocked needles. The blockage in the needles is caused by the presence of adhesive from the assembly process. If the needle is blocked, emergency treatment is delayed which can lead to heart or lung failure, or death. An affected device may cause additional injury since the diagnosis of lung injury may be complicated. If treatment is unsuccessful with the first needle, health care providers may attempt to place another needle and could cause further lung collapse.

There has been one complaint and no reported injuries or deaths.

The TM-317 PneumoDart-Pneumothorax Needle is used to remove air that has become trapped in the pleural cavity during a life threatening situation such as trauma to the lung or a collapsed lung. The needle is used in emergency pre-hospital or hospital settings.

Subscribe to our e-Newsletters
Stay up to date with the latest news, articles, and events. Plus, get special offers
from American Pharmaceutical Review – all delivered right to your inbox! Sign up now!

On November 15, 2019, Tytek Medical sent an Urgent Medical Device Recall Notice to customers and distributors informing them of the affected model and provided the following instructions:

Do not use any of the affected product.

Recall the product from points of distribution and return it right away by UPS Account #R5513V or DHL Account #958855282 to:

Tytek Medical
4700 Ashwood Drive Ste 445
Cincinnati, OH 45241

Complete the Acknowledgement and Receipt Form as soon as possible and return the form by fax to 1-513-874-7294.

Customers who have questions or need additional information about this recall should call 1-513- 872-7326, Monday through Friday between 8:00am and 5:00pm (Eastern Time).

  • <<
  • >>